Kryo 1060
A large capacity controlled rate freezer for higher volume storage.
The system's sample capacity is sufficient for most intense process situations.
A top opening chamber, combined with a unique forced laminar flow pattern of the coolant and its cryogenic insulation, ensures even and accurate temperature control at all phases of the freezing protocol. The stainless-steel finish enables wipe-clean functionality and ruggedness.
Suitable for volume cell line or vaccine storage, this liquid nitrogen CRF has system safety protection with a -100ºC end temperature to ensure sample integrity during transfer to cryostorage.
The new MR7 controller offers password-controlled access on multiple user levels. Updates and alerts are emailed via WiFi to provide peace of mind and unlimited freezer profile storage enabling the operator to select optimum processing conditions.
The Kryo 1060 larger controlled rate freezer range is useful in areas such as:
Biobanks
Cell Based Pharmaceutical
Biotechnology
Medical Researchers
Blood and Tissue Banks
Biorepositories
Translational Medicine
Regenerative Medicine

Key Benefits
Menu driven controller, simple to programme and easy to operate
Unique forced laminar flow pattern allows even and accurate temperature control
Multiple safety features to protect against power and PC failure
Validation offered via controller password-controlled access on multiple user levels
Updates and alerts emailed via Wi-Fi to your PC or phone providing peace of mind
Unlimited* freezer profile storage enabling the operator to select optimum processing conditions
*SD Card has 8GB storage, enough for 6 million profiles or 80,000 Kryofiles.
Principal Features
Chamber Volume: 180 or 380 Litres
Temperature Range: +40°C to -100°C
Heating Rates: 0.01°C to 1°C /min
Cooling Rates: -0.01°C to -5°C /min
System Controller: Planer MR7 Controller
Copes with temporary power/system outs
Time and date stamping
Programme preview before run
High capacity LN2 cylinder: large cooling reservoir - extended hold times